The key component in an RF and microwave calibration system
The Fluke 9640A and 9640A-LPNX RF Reference Sources are designed specifically for RF calibration (including spectrum analyzer calibration, network analyzer calibration, attenuator calibration, and more), featuring a calibration-oriented user interface, precision signal level and attenuation, high signal purity and precision low distortion modulation. This unique combination of features and performance makes them clearly superior to the general purpose signal generators that are often used in RF calibration systems, with the 9604A-LPNX low phase noise version providing superior phase noise performance.
The 9640A Series simplifies and speeds up calibration procedures, reduces opportunities for operator errors, and greatly simplifies RF metrology. As the core of an RF and microwave calibration system, the 9640A Series covers the majority of test points required for calibrating spectrum analyzers of any frequency range.
Used manually or automated with MET/CAL® Plus Calibration Management Software, the 9640A rf reference reduces complexity and calibration times, dramatically improving efficiency and increasing capacity. Walk-away automation with the 9640A Series performing the majority of tests enables operators to carry out additional value-adding tasks rather than wasting time waiting for frequent system setup changes.
What’s new?
New features included in the 9640A Series:
9640A-LPNX model improved phase noise
Reduced close-in phase noise and specifications to 1 Hz offsets.
Improved level accuracy and attenuation
Specifications for better test uncertainty ratios against the most demanding workloads.
Optional integrated 50 MHz frequency counter
Reducing the number of instruments needed in an RF calibration system in the lab or onsite for UUT frequency reference testing.
Extended leveled sine frequency setting resolution
An enhanced mode with 10 uHz resolution at all frequencies, giving a maximum display and setting resolution of 4.023 999 999 999 99 GHz.
Leveled sine minimum output frequency of 1 mHz
Replacing a function generator in many multipurpose calibration systems, enhancing the 9640A Series versatility.
External phase modulation
Also with 1 MHz bandwidth for phase and frequency modulation.
Narrow range-locked sweep
Enhanced resolution frequency sweep with new phase and amplitude continuous, range locked, narrow channel sweep.
New accessory:
9600FLT 1 GHz wide offset phase noise filter
Bandpass filter reducing high offset frequency noise levels for spectrum analyzer calibration 1 GHz phase noise tests.
Summary specifications | ||
|
Frequency specifications |
Level specifications |
Range |
1 mHz to 4 GHz |
-130 to +24 dBm to 125 MHz, |
Resolution |
10 uHz |
0.001 dB |
Accuracy |
0.05 ppm + 5 uHz |
Down to -48 dBm 10 MHz to 128 MHz |
Attenuation |
|
0.02 dB to 55 dB, 0.15 dB at 110 dB |
VSWR |
|
≤500 MHz: ≤1.1, |
Harmonics and spurious |
–60 dBc harmonics, –70 dBc spurious to 1 GHz | |
Phase noise at 1 GHz |
9640A: –122 dBc/Hz, typical, at 10 kHz offset | |
Modulation |
AM, FM, PM, internal and external. Frequency pull and external leveling. | |
Frequency sweep |
1 mHz to 4 GHz. Linear or Logarithmic. Stop-Start or Center-Span. | |
Frequency counter |
Optional internal 50 MHz frequency counter | |
Temperature |
Operating: 0 °C to 50 °C, 23 °C ± 5 °C for specified performance | |
Standard interfaces |
IEEE488.2 (GPIB) | |
GPIB command emulation |
9640A, 9640A-LPNX: HP3335 | |
Calibration documentation |
17025 accredited report of calibration included | |
Dimensions (HxWxD) |
146 mm x 433 mm x and 533 mm (5.8 in x 17.0 in x 21.0 in) Industry-standard 19 in (483 mm) rack mounting when fitted with Y9600 rack mounting kit | |
Weight |
18 kg (40 lbs) |