소개 | 경제적인 비용으로 실험실과 현장에서의 정밀 온도측정 및 교정용 기준을 제공 |
---|---|
AM1751 | - -200°C ~ 670°C
- 정확도 ±0.015°C (@ 0.01°C) - 장기 드리프트 ±0.01°C - 단기 드리프트 0.007°C - 온도계수 0.003925 - W(Ga)>=1.11807 - Inconel™ 소재 - 견고하고 충격에 강함 |
AM1730 | - -200°C ~ 420°C
- 정확도 ±0.015°C (@ 0.01°C) - 장기 드리프트 ±0.01°C - 단기 드리프트 0.007°C - 온도계수 0.003925 - W(Ga)>=1.11807 - Inconel™ 소재 - 견고하고 충격에 강함 |
AM1710 | - -60°C ~ 160°C
- 정확도 ±0.015°C (@ 0.01°C) - 장기 드리프트 ±0.01°C - 단기 드리프트 0.007°C - 온도계수 0.003925 - W(Ga)>=1.11807 - 스테인리스 스틸 소재 - 견고하고 충격에 강함 |
제조사 | 미국 AccuMac Technology |
■ 사양 (AM1751)
Temperature Range |
1751-12: -200 °C to 670 °C
1751-20: -200 °C to 670 °C |
Resistance at 0 °C |
Nominal 100 Ω |
Temperature Coefficient |
0.003925 Ω/ Ω/°C |
Accuracy |
±0.025 °C at -196 °C
±0.015 °C at 0.01 °C
±0.035 °C at 420 °C
±0.05 °C at 661 °C |
Drift |
±0.01 °C at TPW after 100 hours at 661 °C |
Short Term Stability |
±0.007 °C |
Thermal Shock |
±0.005 °C after 10 times thermal cycles from minimum to maximum temperatures |
Hysteresis |
<=0.005 °C |
Self-heating |
50 mW/°C |
Response Time |
9 seconds for 63% response to step change in water moving at 3 feet per second |
Measurement Current |
0.5 mA or 1 mA |
Sensor Length |
32 mm |
Sensor Location |
5 mm from tip |
Insulation Resistance |
>1000 MΩ at room temperature |
Sheath Material |
Inconeltm |
Dimension |
1751-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1751-20: 0.25 inch X 20 inch (6.35 mm X 500 mm) |
External Leads |
Teflontm –insulated copper wire, 4 leads, 2.5 meters |
Handle Dimension |
15 mm (OD) X 65 mm (L) |
Handle Temperature Range* |
-50 °C to 180 °C |
Optional Calibration |
NIST traceable calibration and data available per request, Ordering # 5007 |
*Handle temperature outside this range will cause damage to the probe.
■ 사양 (AM1730)
Temperature Range |
1730-12: -200 °C to 420 °C
1730-9: -60 °C to 300 °C |
Resistance at 0 °C |
Nominal 100 Ω |
Temperature Coefficient |
0.003925 Ω/ Ω/°C |
Accuracy |
±0.025 °C at -196 °C
±0.015 °C at 0.01 °C
±0.035 °C at 420 °C |
Drift |
±0.01 °C at TPW after 100 hours at 420 °C |
Short Term Stability |
±0.007 °C |
Thermal Shock |
±0.005 °C after 10 times thermal cycles from minimum to maximum temperatures |
Hysteresis |
<=0.005 °C |
Self-heating |
50 mW/°C |
Response Time |
9 seconds for 63% response to step change in water moving at 3 feet per second |
Measurement Current |
0.5 mA or 1 mA |
Sensor Length |
32 mm |
Sensor Location |
5 mm from tip |
Insulation Resistance |
>1000 MΩ at room temperature |
Sheath Material |
Inconeltm |
Dimension |
1730-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1730-9: 0.187 inch X 9 inch (4.75 mm X 229 mm) |
External Leads |
Teflontm –insulated copper wire, 4 leads, 2.5 meters |
Handle Dimension |
1730-12: 15 mm (OD) x 65 mm (L)
1730-9: 10 mm (OD) X 50 mm (L) |
Handle Temperature Range* |
-50 °C to 180 °C |
Optional Calibration |
NIST traceable calibration and data available per request:
1730-12: Ordering # 5014
1730-9: Ordering # 5015 |
*Handle temperature outside this range will cause damage to the probe.
■ 사양 (AM1710)
Temperature Range |
-60 °C to 160 °C |
Resistance at 0 °C |
Nominal 100 Ω |
Temperature Coefficient |
0.003925 Ω/ Ω/°C |
Accuracy |
±0.015 °C at 0.01 °C |
Drift |
±0.01 °C at TPW after 100 hours at 160 °C |
Short Term Stability |
±0.007 °C |
Thermal Shock |
±0.005 °C after 10 times thermal cycles from minimum to maximum temperatures |
Hysteresis |
<=0.005 °C |
Self-heating |
50 mW/°C |
Response Time |
9 seconds for 63% response to step change in water moving at 3 feet per second |
Measurement Current |
0.5 mA or 1 mA |
Sensor Length |
32 mm |
Sensor Location |
5 mm from tip |
Insulation Resistance |
>1000 MΩ at room temperature |
Sheath Material |
Stainless Steel |
Dimension |
0.25 inch X 12 inch (6.35 mm X 305 mm) |
External Leads |
Teflontm –insulated copper wire, 4 leads, 2.5 meters |
Handle Dimension |
1710-12: 15 mm (OD) x 65 mm (L)
1710-9: 10 mm (OD) X 50 mm (L) |
Handle Temperature Range* |
-50 °C to 160 °C |
Optional Calibration |
NIST traceable calibration and data available per request:
Ordering # 5017 |
*Handle temperature outside this range will cause damage to the probe.